WIKA operating instructions, model CEP1000

11570351.02 10/2010 GB/D/F/E/I

GB

6. Commissioning, operation

( + )

( - )

LOOPPWR

CEP 1000

MODE

%ERROR

AUTO

MAX

25%

MIN

Calibration Line

SOURCE

mA

6.2 Basic operating modes

The CEP1000 has 4 operating modes for the mA range and one operating mode for

voltage measurement. The following text and figures explain different operating modes.

6.2.1 Milliampere source

Figure 3 shows a typical application, for which the CEP1000 is used as a source of

current for the range from 4 ... 20 mA for the instrument to be tested.

1) Turn on the instrument. As a standard the operating mode source of current in the

mA range is set.

2) Connect the instrument to be calibrated.

For the instrument to function as a source of current, the circuit must be

closed. In case of an interrupted circuit the LCD display "OL" is blinking

and indicating an overload or an interruption of the circuit.

Instrument with an input

of 4 ... 20 mA

Max. load: 1200 Ω without HART™

950 Ω with HART™

Figure 3 - Current source for 4 ... 20 mA

4 ... 20 mA