Commissioning, operation – WIKA CEP1000 User Manual

Page 12

background image

12

WIKA operating instructions, model CEP1000

11570351.02 10/2010 GB/D/F/E/I

GB

5. Transport, packaging and storage / 6. Commissioning ...

Store the instrument in its original packaging in a location that fulfils the conditions

listed above. If the original packaging is not available, pack and store the instrument as

described below:

1. Wrap the instrument in an antistatic plastic film.

2. Place the instrument, along with shock-absorbent material, in the packaging.

3. If stored for a prolonged period of time (more than 30 days), place a bag, containing

a desiccant, inside the packaging.

WARNING!

Before storing the instrument (following operation), remove any residual

media. This is of particular importance if the medium is hazardous to

health, e.g. caustic, toxic, carcinogenic, radioactive, etc.

6. Commissioning, operation

DANGER!

To avoid electric shock and injuries:

Do not connect voltage higher than the nominal voltage. For the

supported voltage ranges refer to "3. Specifications."

Observe all safety measures for the instruments.

Do not use the calibrator if it is damaged. Prior to use check the

housing of the calibrator. Check carefully if plastic parts are missing

and if there are any cracks. Pay special attention to insulation

around connections.

Select the proper function and correct measuring range.

The battery case must be closed and locked before the calibrator is

put into operation.

Disconnect the test leads from the calibrator prior to opening the

battery case.

Check the test leads for signs of damaged insulation or metal parts

with natural finish.

When using the probe keep the fingers away from the probe

contacts. Hold the probes behind the finger protection at the test

leads. Do not use the calibrator in case of malfunction. The instru-

ment protection can be impaired. In case of doubt the instrument

must be repaired.

Do not operate the instrument in the environment with explosive

gases, vapours or dusts.

Disconnect the capillary prior to switching to another measurement

or before using as a simulator.

Use only the specified spare parts for repairs of the calibrator.