WIKA operating instructions, model CEP1000

11570351.02 10/2010 GB/D/F/E/I

GB

6. Commissioning, operation

( + )

( - )

LOOPPWR

CEP 1000

MODE

%ERROR

AUTO

MAX

25%

MIN

Calibration Line

MEASURE

mA

6.2.3 Milliampere measurement

Figure 5 shows the way the CEP1000 must be connected to carry out a measurement in

a circuit for 4 ... 20 mA. During this application the instrument just measures the current

in the circuit and displays it on the main display in mA. The lower (smaller) scale range

indicates the percentage of current for the measuring range 4 ... 20 mA.

6.2.4 Milliampere measurement with 24 V power supply

For this case figure 5 also shows the correct connection. However, in this case a 2-wire

transmitter is tested, it requires an external power supply for the output current. For this

application the instrument supplies the circuit with an operating voltage of 24 V and at

the same time measures the emerging current. To enable the function with separate

power supply, press the "LOOP PWR" key if the option milliampere measurement is

selected.

The measuring mode with separate power supply can be activated for

the option measuring in the milliampere range.

Instrument with an

output for 4 ... 20 mA

4 ... 20 mA

Maximum input signal: 24 mA

To measure the output signal of a two-wire sensor use

the option measurement with separate power supply.

Figure 5: Measurement in the range of 4 ... 20 mA