WIKA operating instructions, model CEP1000

11570351.02 10/2010 GB/D/F/E/I

GB

6. Commissioning, operation

To avoid faulty measurements that can lead to electric shock or

injuries, replace the battery immediately as soon as a dead battery

is indicated.

WARNING!

To avoid damages of the calibrator or devices during testing, use the

correct terminals, functions and measuring ranges for measurement or

simulation.

6.1. First steps

6.1.1 Battery saving function

To avoid discharge of the battery in case of unintended switching on of the CEP1000,

the battery saving function is activated ex works and set to 30 minutes. The battery

saving function turns off the CEP1000 automatically if no key is pressed within 30

minutes.

To change the turn-off time or deactivate it, the following steps must be carried out when

turning on:

1) Turn off the instrument.

2) Press the "AUTO" key.

3) Turn on the instrument again.

4) After turning on wait for 1 second and then release the "AUTO" key.

5) The set period of time appears on the display for the battery saving function

(between "Off" and 30 minutes).

6) Set the required time using the knob (between "Off" and 30 minutes).

7) Press the "AUTO" key again to save the value and switch the instrument back to

normal operation. The saved value is stored in the memory until it is changed by the

operator.