WIKA operating instructions, model CEP1000

11570351.02 10/2010 GB/D/F/E/I

GB

6. Commissioning, operation

6.3 Advanced operating modes

6.3.1 Automatic step function/automatic ramp mode

The CEP1000 is capable of increasing the output current in the mA range in steps of

25 % automatically with intervals of 5 seconds. The output current in the mA range

can be increased linearly in form of the ramp function from 4 mA up to 20 mA and then

reduced again automatically to 4 mA.

To activate the automatic step/ramp function proceed as follows:

1) The automatic step/ramp function can be used for the application as milliampere

source as well as for the milliampere simulation. Switch the instrument into the corre-

sponding operating mode.

2) Connect the CEP1000 to the instrument to be tested.

3) For the automatic step function press the "AUTO" key once, for a slow ramp function

twice and for the quick ramp function three times. After every key stroke a small

confirmation window appears in the lower left corner of the LCD display, it indicates

which step function or ramp function has been selected.

4) As soon as the step function/ramp function has been activated, it is functioning

constantly until another key is pressed.

6.3.2 Fault percentage function

This unique function calculates the fault percentage of the measuring range (a measur-

ing range of 4 ... 20 mA gives a span of 16 mA). Using the function for calculation of fault

percentage the user can "Zero Out" the milliampere indication, thus, enabling indication

of deviation from the ideal value on the display as a fault percentage.

Procedure:

1) Connect the CEP1000 to the instrument to be tested.

2) Set a mode with milliampere output for the instrument (source or simulation).

3) Set the required output value.

4) If the instrument to be tested does not measure the exact value, press and hold

the "%ERROR" key and adjust the knob so that the instrument displays the correct

value. Press and keep holding the “%ERROR” key further on.

5) The indication in the percentage field corresponds to the fault or deviation as a

percentage of the span.

6) After the "%ERROR" key has been released, the indication switches back to normal

operation.