2 connection procedure – WIKA IPT-11 User Manual
Page 14

5
.2 Connection procedure
Proceed as follows:
1
Unscrew the housing cover
2
If an indicating and adjustment module is installed, remove it by
turning it slightly to the left.
3
Loosen compression nut of the cable entry
4
Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5
Insert the cable into the sensor through the cable entry
6
Lift the opening levers of the terminals with a screwdriver (see
following illustration)
7
Insert the wire ends into the open terminals according to the wiring
plan
Fig. 7: Connection steps 6 and 7
8
Press down the opening levers of the terminals, you will hear the
terminal spring closing
9
Check the hold of the wires in the terminals by lightly pulling on
them
10 Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
11 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
14
External indicating and adjustment unit
5 Connecting to power supply
31550
-EN
-100726