Failure

Possible cause

Procedure

Signal span dropping off/too small

Diaphragm is damaged, e.g. through

impact, abrasive/agressive media;

corrosion of diaphragm/pressure

connector; transmission fluid missing.

Contact the manufacturer and replace

the instrument

Signal span drops off

Moisture present (e.g. at the cable tail) Install the cable correctly

Signal span erratic

Working temperature too high/too low Ensure permissible temperatures as per

the Operating Instructions

Signal span incorrect

Working temperature too high/too low Ensure permissible temperatures as per

the Operating Instructions

Abnormal zero point signal

Medium or ambient temperature too

high/too low

Control the internal temperature of the

instrument within the permissible range;

observe the allowable temperature error

(see Operating Instructions)

Diaphragm is damaged, e.g. through

impact, abrasive/agressive media;

corrosion of diaphragm/pressure

connector.

Replace instrument

Abnormal zero point signal

Working temperature too high/too low Ensure permissible temperatures as per

the Operating Instructions

Zero point signal unstable/ too low/high Moisture present (e.g. at the cable tail) Install the cable correctly,

filter element insertion

Failure

Possible cause

Procedure

Hot instrument case surface

Permissible ambient or medium tempe-

rature exceeded

Ensure permissible ambient/medium

temperature limits are observed (see

Operating Instructions)

9. Trouble shooting

GB

Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the
right voltage supply and the right type of wiring (2-wire) has been chosen?

9. Trouble shooting / 10. Storage, disposal

GB

If the problem persists, contact our sales department.
USA, Canada
If the problem continues, contact WIKA or an authorized agent for assistance. If the level
probe must be returned obtain an RMA (return material authorization) number and shipping
instructions from the place of purchase. Be sure to include detailed information about the
problem. Level probes received by WIKA without a valid RMA number will not be accepted.
Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.wika.de / www.wika.com).

In case of unjustified reclamation we charge the reclamation handling expenses.
*) Make sure that after the assembly the unit is working properly. In case the error continues to exist send in the instrument for

reparation (or replace the unit).

10. Storage, disposal

When storing or disposing of the level probe, take precautions with regard
to remaining media in removed level probes. We recommend cleaning the
level probe properly and carefully. Remaining media in the pressure port
may be hazardous or toxic!

Dispose of instrument components and packaging materials in accordance with the
respective waste treatment and disposal regulations of the region or country to which
the instrument is supplied.

Disposal

Mount the protection cap when storing the pressure transmitter in order to prevent any
damage to the diaphragm.

Storage

Take precautions with regard to remaining media in removed level probes.
Remaining media in the pressure port may be hazardous or toxic!
Remove the level probe from service and mark it to prevent it from being
used again accidentally, if it becomes damaged or unsafe for operation.
Have repairs performed by the manufacturer only.

Do not insert any pointed or hard objects into the pressure port for cleaning to prevent
damage to the diaphragm.

. Trouble shooting

!

Warning

!

Warning