5 measurement setup density and interface – WIKA DPT-10 User Manual
Page 23
+
–
min.
max.
1
2
3
4
5
+
–
Fig. 19: Measurement setup in closed vessel with superimposed steam
1
Condensate vessel
2
Blocking valve
3
Precipitator
4
Drain valve
5
DPT10
4
.5 Measurement setup density and interface
Density and interface measurement are carried out in level measure-
ment mode.
à Mount DPT10 below the lower isolating diaphragm
à The ambient temperature should be the same for both capillaries
Information:
The density measurement is only ensured at a level above the upper
measuring point. Homogeneous density allocation is a prerequisite.
=
∆p
h g
+
–
0
,3
m
Fig. 20: Measurement setup with density measurement. Density 1 and h = 0.3 m
means Δp = 29.4 mbar. With density change: e.g. measured Δp = 35.3 mbar
corresponds to density 1.2, measured Δp = 23.5 mbar corresponds to density
0
.8
Density measurement
Differential pressure transmitter DPT10 - metal measuring diaphragm • Foundation Fieldbus
23
4 Mounting
37246
-
EN
-100120
- 890.09.2190 (44 pages)
- A-10 (96 pages)
- A2G-50 (52 pages)
- A2G-55 (36 pages)
- AC-1 (88 pages)
- C-2 (24 pages)
- D-10-7 (112 pages)
- D-20-9 (51 pages)
- D-20-9 (35 pages)
- DG-10 (112 pages)
- DP-10 (44 pages)
- DPT-10 (96 pages)
- GCS-1 (76 pages)
- HP-2 (84 pages)
- IL-10 (31 pages)
- IPT-10 (28 pages)
- IPT-10 (36 pages)
- IPT-10 (56 pages)
- IPT-10 (48 pages)
- IS-20-F (7 pages)
- IS-20-F (43 pages)
- IS-20-H (7 pages)
- UT-10 (42 pages)
- UT-10 (52 pages)
- IUT-10 (78 pages)
- LH-10 (60 pages)
- LH-20 (60 pages)
- LS-10 (60 pages)
- MG-1 (92 pages)
- MH-1 (11 pages)
- MH-2 (9 pages)
- MHC-1 (84 pages)
- N-10 (35 pages)
- O-10 (108 pages)
- OT-1 (60 pages)
- P-30 (92 pages)
- PSA-31 (124 pages)
- PSD-30 (128 pages)
- R-1 (92 pages)
- S-10 (35 pages)
- S-11 (102 pages)
- S-20 (52 pages)
- S-20 (96 pages)
- SL-1 (21 pages)