beautypg.com

Dynalab SMP30 User Manual

Page 7

background image

5

5. To begin the melt cycle, use the upper menu

key next to “Start Ramp” to instruct the
instrument to commence the melt program.

6. Once the melt program commences the

display will change to show the
pre-programmed ramping rate and the
temperature displayed on the screen will
increase at the rate selected in the method.

7. Once the melt determination is complete use

the Enter key to select “DONE” to end the
instruments heating program.

5.5 Saving Melt Temperatures

1. During the ramping phase of the melt

determination the I, II and III keys can be used
to record up to seven temperatures for each
capillary.

2. An audible bleep will sound each time these

keys are pressed and the recorded
temperature will be displayed on the screen
above the button. Each additional time the
key is pressed another audible bleep will be
heard and the temperature displayed will
update to the last recorded temperature.

Note:-. If an eighth attempt to record a value
is attempted, a longer audible bleep
will
sound to indicate that the memory location
is full.

5.6 Viewing Stored Temperatures

1. Once a melt determination is complete and

the Enter key has been pressed to select
“DONE”, the following screen will be
displayed:

2. Use the upper menu key next to “PRINT

RESULTS” to instruct the instrument to print
the stored temperature values on theto the
accessory printer.

Note:- To be able to print the stored
temperature values the accessory printer
SMP30/1 needs to be connected to the
instrument.

3. Use the lower menu key next to “VIEW

RESULTS” to display the results screen as
shown below:

Note:- A maximum of two results will be
displayed on the screen at any one time.

4. Use the upper and lower menu keys next to

“UP” and “DOWN” to scroll through the
recorded results for each tube. Alternatively,
an individual tube’s results can be displayed
using the corresponding I, II and III keys and
once again, by using the upper and lower
menu keys next to “UP” and “DOWN” to
navigate through the recorded results.

5. Once the results have been reviewed use the

Enter key to select “DONE” to return to the
screen in step 1.

09:00

112.0°C

RAMPING AT 1.0°C/min

DONE 108.3°C 109.5°C 109.2°C

09:00

112.0°C

PRINT

RESULTS

VIEW

RESULTS

EXIT

TUBE 1 RESULT 1-2 OF 7

112.0°C

TEMPERATURE 106.2°C 108.2°C

UP

RESULT TIME 09:00 09:02

DOWN

DONE

09:00

112.0°C

RAMPING AT 1.0°C/min

DONE