Melting point apparatus smp30 instructions for use, Introduction, Safety advice before use – Dynalab SMP30 User Manual
Page 3: General description

1
Melting Point Apparatus
SMP30
Instructions for use
1. Introduction
Thank you for purchasing this piece of Stuart 
equipment. To get the best performance from 
the SMP30 please read these instructions 
carefully before use. Before discarding the 
packaging check that all parts are present and 
correct.
For your own safety and that of others 
please read and understand the safety 
advice given below before using the 
equipment.
2. Safety advice before use
If the equipment is not used in the manner 
described in this manual and with 
accessories other than those recommended by 
Stuart, the protection provided might be 
impaired.
This equipment is designed to operate under the 
following conditions:
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuation not exceeding 10%
❖ Over-voltage category II IEC 60364-4-443
❖ Pollution degree 2
Never move or carry the unit when in use or 
connected to the mains electricity supply.
 
The unit should be carried using both hands with 
fingers under each side of the frame.
The ventilation slots at the back of the sample 
block should not be sealed or obstructed.
3. General description
The apparatus is designed to measure and record 
the temperatures of crystalline samples held 
within capillary tubes. Up to three tubes are 
accommodated in an illuminated chamber within 
the aluminium block. The tubes are viewed 
through a magnifying lens on the front of the 
unit. All controls are via the membrane keypad.
The temperature range is ambient to 400°C and 
the heating rate is variable between 0.5°C per 
minute and 10°C per minute. Temperatures and 
instrument information are displayed on the LCD 
screen. Up to seven temperatures per capillary 
tube can be saved and recalled on the screen. 
Three methods can be stored for repeated use 
and the plateau temperature and heating rate 
are fully programmable. When the melt cycle is 
started the SMP30 heats up to the 
pre-programmed plateau temperature at the 
maximum rate before stabilising at the plateau 
temperature for 120 seconds. An alarm will 
sound to indicate that the SMP30 is ready to 
start to heat at the pre-programmed rate. Once 
the melt has occurred and the cycle stopped, 
cooling is automatic to ambient temperature.
The SMP30 has English, Spanish, French, Italian 
and German language options.