Myron L 6Psi and 4P User Manual
Page 33
29
6.
Press
key to accept the setting. Pressing any of the
measurement keys will exit without changing the setting. “User”
mode “Linc” is now complete. The “User” mode will now use the
calibration gain constant used for the calibration of the
Standard Solution as outlined above.
C. Canceling “User” Mode Calibration “Linc”
The Ultrameter II must be in “User” linked mode in order to cancel the
“Linc” (ref. SOLUTION SELECTION, pg. 11).
1.
Press “Linked” measurement key
,
or .
Two solution icons will be shown in the left side of display -
“User” and another, e.g., “KCl”.
2.
Press
key until the menu “Linc” appears (see Figure 27).
3.
Press
key; the instrument will display both “SEL” and
the “User” Icon.
4.
Press
key until “User” is the only solution icon being
displayed.
5.
Press
key.
6.
The “User” mode calibration “Linc” has now been canceled.
NOTES:
1. To maintain repeatability, use the same standard solutions for future
calibrations.
2. Calibration of the Ultrameter II Gain Factor for “User” mode is not
available when the calibration linc has been established. The other
calibration functions (i.e., Temperature Compensation %/C settings and
TDS Ratio settings) are still intact. To perform a calibration of the “User”
mode as described in User Calibration Conductivity/TDS, pg. 15, the
“User” mode Linc should be canceled. See above, “Canceling “User”
mode calibration “Linc””.
3. Once a “Linc” has been established for “User” mode , the “Linc” will
apply to all measurement modes using User solution selection (i.e.,
TDS/User, Cond/User or Res/User).