Myron L 6Psi and 4P User Manual
Page 20
16
a.
Rinse conductivity cell three times with your standard.
b.
Refill conductivity cell with same standard.
c.
Press
or
, then press
twice in COND/three
times in TDS. The “CAL” icon will appear on the display.
d.
Press
or
to step the displayed value toward the
standard’s value or hold a key down to scroll rapidly through
the reading.
e.
Press
once to confirm new value and end the
calibration sequence for this particular solution type.
3. Resistivity Calibration
Resistivity is the reciprocal of Conductivity. To calibrate resistivity,
calibrate conductivity for the solution type you wish to measure (ref.
Conductivity or TDS Calibration, pg. 15).
4. Reloading Factory Calibration (Cond or TDS)
If calibration is suspect or known to be incorrect, and no standard solution
is available, the calibration value can be replaced with the original factory
value for that solution. This “FAC” value is the same for all Ultrameter IIs,
and returns you to a known state without solution in the cell. The “FAC”
internal electronics calibration (which bypasses the electrodes and cell)
is not intended to replace calibration with conductivity/TDS standard
solutions. If another solution type requires resetting, change solution
type and repeat this procedure.
a.
Press
or
.
b.
Press
. (If in “User” solution mode, press CAL key
twice if in Conductivity, and three times if in TDS to skip over
tempco and ratio adjustments.)
c.
Press
key until “FAC” appears and release.
d.
Press
to accept the factory calibration setting.