Myron L 6Psi and 4P User Manual
Page 32

28
4.
Verify/Set the calibration linc. (See below – Setting “User”
Mode Calibration “Linc”.)
B. Setting “User” mode Calibration “Linc”
The Linc function sets or “links” the calibration gain factor of a Standard
Solution to the User solution mode. Once set, the “Linc” will stay intact
with future calibrations unless the Linc has been canceled. For more
information on canceling the “User” mode Calibration Linc refer to the
section “Canceling “User” mode Calibration “Linc””, pg. 29.
Follow the steps below to set either the KCl, NaCl or 442 calibration
factor to the User solution mode.
1.
Press measurement key desired to be “Linked”, i.e.,
,
or
.
2.
Place the Ultrameter II in
“User” mode (ref. SOLUTION
SELECTION, pg. 11, for
selecting the “User” mode).
3.
Press
arrow key until
the menu “Linc” appears
(see Figure 27).
4.
Press
key. The
instrument will display “SEL”
and the “User” Icon (see
Figure 28).
Any additional display of KCl,
NaCl or 442 icons indicates a “Linc”
between the User solution and the
other solution displayed.
5.
Press
or
keys
to select a Standard
Solution to be linked to the
“User” mode calibration
constant (see Figure 29). “User” is linked to “KCl”.
If none of the Solution Selection icons are displayed (i.e., KCl, NaCl or
442), nothing has been linked to “User” mode.
Figure 27
Figure 28
User
Figure 29
KCl
User