Myron L 6Psi and 4P User Manual
Page 31

27
2.
Press
to display the stored memory records.
3.
Press
repeatedly until you pass the CELL ch location.
The display will show “Auto oFF” (see Figure 24).
4.
Press
to initiate. “CAL”
will be displayed along with
“15 SEC” or current Auto
oFF value (see Figure 25).
5.
Press
or
to
change the amount of time
(see Figure 26). Maximum
time is shown.
6.
Press
to accept the
change (new time).
XV.
“USER” MODE CALIBRATION LINC™ FUNCTION
The Linc™ function allows easy calibration when in “User” mode and the
user does not have a user standard solution to calibrate the instrument.
This function will ensure more repeatable and accurate measurements
than many other calibration methods. It is recommended that this
function be used to provide the highest degree of confidence when the
Ultrameter II is used in “User” mode. When Linc is used, the “User”
mode is linked to another standard, i.e., if User and KCl are linked, a KCI
standard solution is used to calibrate the instrument. It is that simple.
A. Calibration of Ultrameter II for use in “User” mode
1.
Press
or
key.
2.
Calibrate the unit using a Standard Solution (ref.
CALIBRATION, pg. 14).
3.
Place the Ultrameter II in “User” mode (ref. SOLUTION
SELECTION, pg. 11).
Figure 25
CAL
Figure 26
CAL