N o t e – Agilent Technologies G1978B User Manual
Page 40
![background image](/manuals/44748/40/background.png)
40
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
3
Methods
To create a method for alternating ESI and APCI operation
9 Click the OK button.
Figure 23
MM-ES and MM-APCI switching using Set Up MSD Signals dialog box
N O T E
In general, use the mixed mode operation (MM-ES+APCI setting with Signal 1) instead of
alternating MM-ES (Signal 1) and MM-APCI (Signal 2) mode switching. Two times as many
scans are obtained during elution of a chromatographic peak, and no delay is needed
between scans, resulting in better data. It is rarely necessary to know whether a compound
responds purely in ESI or APCI modes on a chromatographic time scale.
Ionization mode for
Signal 1 and Signal 2