Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

31

Verification

2

To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation

d Click OK.

8 Repeat for the MMCheckSL_CI.M data file. Use the same subdirectory and

change Data File to Multi_CI.

9 Repeat for the MMCheckSL_MX.M data file. Use the same subdirectory and

change Data File to Multi_MX as the data file name.

10 Click Run to start the sequence.

11 Review the results.

When the last method has finished, the report “Multimode Verification Report”
prints.