Scan modes – Agilent Technologies G1978B User Manual

Page 27

background image

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

27

Verification

2

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode,
Positive Scan Modes

Preparation

1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric

flask. Use a clean graduated pipette.

2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium

formate.

3 Transfer 5 mL of the first dilution to a 50 mL volumetric flask. Use a clean

graduated pipette.

4 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium

formate.

This provides the final 10 pg/µL reserpine concentration required for
performance verification.

5 Transfer approximately 1 mL of the second dilution to an autosampler vial.

Table 3

6130 and 6140 Quad Performance Verification Summary, Scan Mode

G1978B
ES Positive Scan Mode

G1978B
APCI Positive Scan Mode

Sample

Reserpine, 5 ng/µL

Reserpine, 5 ng/µL

Concentration after dilution

10 pg/µL

10 pg/µL

Injection volume

5 µL

5 µL

Total sample amount injected

50 pg

50 pg

Sample order number

G2423A

G2423A

Solvent

75:25
methanol/water

with 5mM

ammonium formate

75:25
methanol/water

with 5mM

ammonium formate