Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

2

Verification

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes

When the method is finished, a report prints showing the signal-to-noise ratio
for the five blank and five sample peaks, and a blank-subtracted average of the
sample peaks. The blank-subtracted average signal-to-noise value must be
greater than or equal to 20:1 peak-to-peak (100:1 RMS).

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode,
Positive SIM Modes

Preparation

1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric

flask. Use a clean graduated pipette.

2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium

formate.

3 Transfer 1mL of the first dilution to a 100 mL volumetric flask. Use a clean

graduated pipette.

4 Dilute to the 100 mL mark with 75:25 methanol/water with 5 mM

ammonium formate.

This provides the final 1 pg/µL reserpine concentration required for perfor-
mance verification.

5 Transfer approximately 1 mL of the second dilution to an autosampler vial.

Table 2

6130 and 6140 Quad Performance Verification Summary, SIM Mode

G1978B
ES Positive SIM Mode

G1978B
APCI Positive SIM Mode

Sample

Reserpine, 5 ng/µL

Reserpine, 5 ng/µL

Concentration after dilution

1 pg/µL

1 pg/µL

Injection volume

1 µL

1 µL

Total sample amount injected

1 pg

1 pg

Sample order number

G2423A

G2423A