N o t e – Agilent Technologies G1978B User Manual
Page 17
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
17
Installation
1
To convert from ESI, APCI or APPI to the multimode source
Figure 13
Multimode spray shield
13 Screw the multimode spray shield into the holder for the spray shield. See
Figure 14
.
Figure 14
Multimode spray shield installed
14 Remove the shipping cover from the multimode source spray chamber.
Field shaping electrode
9 o'clock position
Field shaping electrode
6 o'clock position
N O T E
The field shaping electrodes should be in the nine o’clock and the six o’clock position.
Loosen the end plate screws on each side to adjust the field shaping electrodes position.