Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

17

Installation

1

To convert from ESI, APCI or APPI to the multimode source

Figure 13

Multimode spray shield

13 Screw the multimode spray shield into the holder for the spray shield. See

Figure 14

.

Figure 14

Multimode spray shield installed

14 Remove the shipping cover from the multimode source spray chamber.

Field shaping electrode
9 o'clock position

Field shaping electrode
6 o'clock position

N O T E

The field shaping electrodes should be in the nine o’clock and the six o’clock position.
Loosen the end plate screws on each side to adjust the field shaping electrodes position.