WindSonic Two-Dimensional Sonic Anemometer
press the
SDI-12 address.
• To change the SDI-12 address, press the
prompt enter the command “SDI12>aAb!”; where a is the current address
from the above step and b is the new address. The WindSonic4 will
change its address and the datalogger will exit the SDI-12 Transparent
Mode and respond with “Fail”.
• Verify the new SDI-12 address. Press the
“CR200(X)>” prompt enter the command “SDI12>?!” and press the
FIGURE 7-3. SDI-12 Transparent Mode on a CRBasic CR200(X)-
series datalogger using control port C1/SDI12 and changing the
SDI-12 address from 0 to 1
8. Maintenance
There are no user-serviceable parts on the WindSonic. Keep the transducer
paths clear of any obstructions. When clearing the transducer paths, do not
remove or damage the transducer matching layer. The transducers can be
gently cleaned with a cloth and mild detergent. Do no use solvents and avoid
scratching or damaging the matching layers. The transducer’s matching layers
are the “rubber” caps on each of the transducers. Should the WindSonic be
damaged, fail to output data, or send a nonzero diagnostic, return it for repair