beautypg.com

2 equipment and materials, 3 procedure for sediment, Equipment and materials – Campbell Scientific OBS-5+ Sediment Concentration Monitor and Pressure Sensor User Manual

Page 32: Procedure for sediment

background image

OBS-5+ System

Sediment concentrations can also be calculated manually with the following
equations:

+

=

s

s

i

s

M

V

M

l

/

g

ρ

g/l)

103

x

2.65

(usually

density

ediment

S

(liters)

volume

nitial

I

V

(g)

mass

Sediment

M

s

i

s

=

=

=

ρ

7.1.2 Equipment and Materials

• Dry, completely disaggregated bottom sediment or suspended matter from

the monitoring site

• Large black, neoprene or polyethylene tub

• 1-gallon (4 l) brown Nalgene polypropylene bottle with top cut off

• 1-liter volumetric flask

• Hand-drill motor (manual suspender)

• Mixing propeller (manual suspender)

• Scale with 10 mg accuracy

• Automatic suspender (optional)

• Tea cup with round bottom and teaspoon

7.1.3 Procedure for Sediment

1) Put batteries in the OBS-5+ and connect it to a PC with test cable using the

RS-232 plug (FIGURE 6-1).

2) Start OBS-5+ Utility program; wake the OBS-5+; and click the

OBS

Settings button to verify its response.

24