5. Programming

5.1 Programming Overview

Typical datalogger programs to measure the CS547 consist of four parts:

1.

Measurement of EC and temperature

2.

Correction of ionization errors in EC measurements

3.

Correction of temperature errors in EC measurements

4.

Output processing

All example programs may require modification by the user to fit the specific
application's wiring and programming needs. All example programs in this
manual are for the CR10(X) and assume that datalogger wire connections are as
follows: the LO COND lead is connected to 1L, the HI COND to 1H, the EX
COND to EX1, the EX TEMP to EX2, and the SE TEMP to SE3.

5.2 Measurement Programming.

EC: Results from Instructions 5 or 6 (chosen automatically as part of the
autoranging feature of the following program segment) are processed with
Instruction P59 to produce the resistance across the electrodes:

Input Location Labels

Definitions for the following program:
Rs Solution

resistance

Rp

Resistance of leads/cable and blocking caps

Ct

Solution EC with no temp. correction

C25mScm_1

EC corrected for temperature

AG

SE3

EX2

EX1

1H

1L

Clear (Shield)

Red (Temp)

Orange (Cond)

Black (Ex Cond)

Green (Ex Temp)