2 temperature measurement error – Campbell Scientific CS547 Conductivity and Temperature Probe and A547 Interface User Manual
Page 18

CS547 Conductivity and Temperature Probe and A547 Interface
14
8.2 Temperature Measurement Error
The overall probe accuracy is a combination of the thermistor's interchangeability
specification, the precision of the bridge resistors, and the polynomial error. In a
"worst case" all errors add to an accuracy of
±
0.4
°C over the range of -24° to 48°C
and
±0.9°C over the range of -38°C to 53°C. The major error component is the
interchangeability specification of the thermistor, tabulated in Table 8.2-1. For the
range of 0
° to 50°C the interchangeability error is predominantly offset and can be
determined with a single point calibration. Compensation can then be done with an
offset entered in the measurement instruction. The bridge resistors are 0.1%
tolerance with a 10 ppm temperature coefficient. Polynomial errors are tabulated in
Table 8.2-2 and plotted in Figure 8.2-1.
TABLE 8.2-1. Thermistor
Interchangeability Specification
Temperature
Temperature (
°C)
Tolerance (
±°C)
−40
0.40
−30
0.40
−20
0.32
−10
0.25
0 to +50
0.20
TABLE 8.2-2. Polynomial Error
-40 to +56
<
±1.0°C
-38 to +53
<
±0.5°C
-24 to +48
<
±0.1°C
FIGURE 8.2-1. Error Produced by Polynomial Fit to Published Values