8.2 Temperature Measurement Error

The overall probe accuracy is a combination of the thermistor's interchangeability
specification, the precision of the bridge resistors, and the polynomial error. In a
"worst case" all errors add to an accuracy of

±

0.4

°C over the range of -24° to 48°C

and

±0.9°C over the range of -38°C to 53°C. The major error component is the

interchangeability specification of the thermistor, tabulated in Table 8.2-1. For the
range of 0

° to 50°C the interchangeability error is predominantly offset and can be

determined with a single point calibration. Compensation can then be done with an
offset entered in the measurement instruction. The bridge resistors are 0.1%
tolerance with a 10 ppm temperature coefficient. Polynomial errors are tabulated in
Table 8.2-2 and plotted in Figure 8.2-1.

TABLE 8.2-1. Thermistor

Interchangeability Specification

Temperature

Temperature (

°C)

Tolerance (

±°C)

−40

0.40

−30

0.40

−20

0.32

−10

0.25

0 to +50

0.20

TABLE 8.2-2. Polynomial Error

-40 to +56

<

±1.0°C

-38 to +53

<

±0.5°C

-24 to +48

<

±0.1°C

FIGURE 8.2-1. Error Produced by Polynomial Fit to Published Values