CS547 Conductivity and Temperature Probe and A547 Interface

17

Example 2. Sample CR10(X) Instructions Using AC Half Bridge

01: AC Half Bridge (P5)

01:

1

Rep

02:

22**

7.5 mV 60 Hz rejection Range

03:

9*

IN Chan

04:

3*

Excite all reps w/EXchan 3

05:

2000**

mV Excitation

06:

11*

Loc [:Air_Temp ]

07:

800

Mult

08:

0

Offset

02: Polynomial (P55)

01:

1

Rep

02:

11*

X Loc Air_Temp

03:

11*

F(X) Loc [:Air_Temp ]

04:

-53.46

C0

05:

90.807

C1

06:

-83.257

C2

07:

52.283

C3

08:

-16.723

C4

09:

2.211

C5

* Proper entries will vary with program and datalogger channel and input location

assignments.

** On the 21X and CR7 use the 15 mV input range and 4000 mV excitation.

12. Long Lead Lengths Temperature

If the CS547 has lead lengths of more than 300 feet, use the DC Half Bridge
instruction (Instruction 4) with a 2 millisecond delay to measure temperature.
The delay provides a longer settling time before the measurement is made. Do
not use the CS547 with long lead lengths in an electrically noisy environment.

Example 3. Sample Program CR10 Using DC Half Bridge with Delay

01: Excite, Delay,Volt(SE) (P4)

01:

1

Rep

02:

2**

7.5 mV slow range

03:

9*

IN Chan

04:

3*

Excite all reps w/EXchan 3

05:

2

Delay (units .01sec)

06:

2000**

mV Excitation

07:

11*

Loc [:Temp_C ]

08:

.4***

Mult

09:

0

Offset

(cont.)