beautypg.com

Maintenance, Analysis of errors, 1 ec measurement error – Campbell Scientific CS547 Conductivity and Temperature Probe and A547 Interface User Manual

Page 16

background image

CS547 Conductivity and Temperature Probe and A547 Interface

12

09: Z=X+F (P34)

01:

2

X Loc t

02:

-25

F

03:

3

Z Loc [(t-25).01]

10: Z=X*F (P37)

01:

3

X Loc (t-25).01

02:

.01

F

03:

3

Z Loc [(t-25).01]

11: Polynomial (P55)

01:

1

Rep

02:

3

X Loc (t-25).01

03:

4

F(X) Loc [f(t) ]

04:

.99124 C0

05:

-1.8817

C1

06:

3.4789

C2

07:

-3.51

C3

08:

-1.2

C4

09:

-43

C5

12: Z=1/X (P42)

01:

4

X Loc f(t)

02:

6

Z Loc [1/f(t) ]

13: Z=X*F (P37)

01:

6

X Loc 1/f(t)

02:

1.408

F

EC of calibration solution

03:

7

Z Loc [Act'l Con]

14: Z=X*Y (P36)

01:

7

X Loc Act'l Con

02:

1

Y Loc Rs

03:

8

Z Loc [Kc (cm-1)]

End

7. Maintenance

Routine maintenance includes thoroughly cleaning the orifice of the CS547
probe with the black nylon brush provided and a little soapy water. Rinse
thoroughly.

8. Analysis of Errors

8.1 EC Measurement Error

1.

Bridge Measurement Error: < 1.0%

2.

Calibration Error:
bridge measurement: < 0.5%
calibration solution: < 1.0%

3.

Ionization Error of KCl and Na+ Solutions After Correction:
< 2.0%, 0.45 to 7.0 mS cm

-1

< 8.0%, 0.005 to 0.45 mS cm

-1