CS547 Conductivity and Temperature Probe and A547 Interface

11

3.

Connect the CS547 and A547 or probe and interface to the datalogger
using the wiring described in Section 4. Enter the following program into
the datalogger.

The calibration solution temperature must be between 1°C and 35°C; the
polynomial in step 11 (P58) corrects for temperature errors within this range.
The solution constant of 1.408 mS cm

-1

(for prepared solution mentioned above),

entered in step 13 (P37), is valid only for a 0.01 molal KCl solution. Location 8
[Kc (cm-1)], generated by step 14, will contain the resultant cell constant.

01: AC Half Bridge (P5)

01:

1

Rep

02:

15

2500 mV fast Range (5000 mV fast for 21X)

03:

2

IN Chan

04:

1

Excite all reps w/EXchan 1

05:

2500

mV Excitation (5000 mV for 21X)

06:

1

Loc [Rs ]

07:

1

Mult

08:

0

Offset

02: BR Transform Rf[X/(1-X)] (P59)

01:

1

Rep

02:

1

Loc [Rs ]

03:

1

Multiplier (Rf)

03: Z=F (P30)

01:

nnn

F

Enter Cable Length in Feet

02:

00

Exponent of 10

03:

5

Loc [Rp ]

04: Z=X*F (P37)

01:

5

Loc Rp

02:

.00032 F

03:

5

Loc [Rp ]

05: Z=X*F (P37)

01:

5

Loc Rp

02:

-.1

F

03:

5

Loc [Rp ]

06: Z=X+F (P34)

01:

5

Loc [Rp ]

02:

-.005

03:

5

Loc [Rp ]

07: Z=X+Y (P33)

01:

1

X Loc Rs

02:

5

Y Loc Rp

03:

1

Z Loc [Rs ]

08: Temp 107 Probe (P11)

01:

1

Rep

02:

3

IN Chan

03:

2

Excite all reps w/EXchan 2

04:

2

Loc [t]

05:

1

Mult

06:

0

Offset

(cont.)