18

02: Polynomial (P55)

01:

1

Rep

02:

11*

X Loc Temp_C

03:

11*

F(X) Loc [:Temp_C ]

04:

-53.46

C0

05:

90.807

C1

06:

-83.257

C2

07:

52.283

C3

08:

-16.723

C4

09:

2.211

C5

*

Proper entries will vary with program and datalogger channel and input location
assignments.

**

On the 21X and CR7 use the 15 mV input range and 4000 mV excitation.

*** Use a multiplier of 0.2 with a 21X and CR7.

13. CS547 Schematic

Black (Ex Cond)

Green (Ex Temp)

Red (Temp)

Orange (Cond)

FIGURE 13-1. CS547 Conductivity and Temperature Circuit Diagram