CS547A Conductivity and Temperature Probe and A547 Interface

Example 12-1. CR1000 measurement instruction with 20 mSec
(20000 uSec) delay:

Therm107(TempDeg_C,1,3,2,20000,_60Hz,1.0,0.0)

Example 12-2. Sample Program CR10 Using DC Half Bridge with Delay

1: Excite, Delay,Volt(SE) (P4)
1: 1

Rep

2: 2**

7.5 mV slow range

3: 3*

IN

Chan

4: 2*

Excite all reps w/EXchan 2

5: 2

Delay

(units

.01sec)

6: 2000** mV

Excitation

7: 11*

Loc [ Temp_C ]

8: .4***

Mult

9: 0

Offset


2: Polynomial (P55)
1: 1

Rep

2: 11*

X Loc Temp_C

3: 11*

F(X) Loc [ Temp_C ]

4: -53.46 C0
5: 90.807 C1
6: -83.257 C2
7: 52.283 C3
8: -16.723 C4
9: 2.211

C5


*

Proper entries will vary with program and datalogger channel and input location assignments.

** On the 21X and CR7 use the 15 mV input range and 4000 mV excitation.
*** Use a multiplier of 0.2 with a 21X and CR7.

13. CS547A Schematic

Black (Ex Cond)

Green (Ex Temp)

Red (Temp)

Orange (Cond)

FIGURE 13-1. CS547A Conductivity and Temperature circuit diagram