beautypg.com

Campbell Scientific CS547A-L Conductivity/Temperature Probe and A547A Interface User Manual

Page 15

background image

CS547A Conductivity and Temperature Probe and A547 Interface

;
;Subtract resistance errors (Rp) caused by the blocking capacitors
;(0.005Kohm) and the cable length (0.000032kohm/ft). Enter cable lead
;length in nnn below.
;

17: Z=F (P30)
1: nnn

F

Enter cable length in feet.

2: 00

Exponent

of

10

3: 5

Z Loc [ Rp ]


18: Z=X*F (P37)
1: 5

X Loc [ Rp ]

2: .00032 F
3: 5

Z Loc [ Rp ]


19: Z=X*F (P37)
1: 5

X Loc [ Rp ]

2: -.1

F

3: 5

Z Loc [ Rp ]


20: Z=X+F (P34)
1: 5

X Loc [ Rp ]

2: -.005

F

3: 5

Z Loc [ Rp ]


21: Z=X+Y (P33)
1: 1

X Loc [ Rs ]

2: 5

Y Loc [ Rp ]

3: 1

Z Loc [ Rs ]


;EC is then calculated by multiplying the reciprocal of resistance,
;which is conductance, by the cell constant.

NOTE: The cell constant (Kc) is printed on the label of each sensor or it can be calculated (see
Section 6.4). It is entered in place of nnn below.

22: Z=1/X (P42)
1: 1

X Loc [ Rs ]

2: 2

Z Loc [ one_ovrRs ]


23: Z=X*F (P37)
1: 2

X Loc [ one_ovrRs ]

2: nnn

F

Enter cell constant.

3: 3

Z Loc [ Ct ]


;
;The following program set corrects for errors of ionization in the EC
;measurement.
;

9