CS547A Conductivity and Temperature Probe and A547 Interface

11: Polynomial (P55)
1: 1

Rep

2: 3

X

Loc

[T25_01]

3: 4

F(X) Loc [f_of_T ]

4: .99124 C0

5: -1.8817 C1

6: 3.4789 C2

7: -3.51

C3

8: -1.2

C4

9: -43

C5


12: Z=1/X (P42)
1: 4

X Loc [f_of_T ]

2: 6

Z Loc [one_ovrfT ]


13: Z=X*F (P37)
1: 6

X Loc [one_ovrfT ]

2: 1.408

F

EC of calibration solution

3: 7

Z

Loc

[Conductiv]


14: Z=X*Y (P36)
1: 7

X

Loc

[Conductiv]

2: 1

Y Loc [Rs ]

3: 8

Z Loc [Kc ]


End

7. Maintenance

Routine maintenance includes thoroughly cleaning the orifice of the CS547A
probe with the black nylon brush provided and a little soapy water. Rinse
thoroughly.

8. Analysis of Errors

8.1 EC Measurement Error

1. Bridge Measurement Error: < 1.0%

2. Calibration

Error:

bridge measurement: < 0.5%
calibration solution: < 1.0%

3. Ionization Error of KCl and Na+ Solutions After Correction:

< 2.0%, 0.45 to 7.0 mS cm

-1

< 8.0%, 0.005 to 0.45 mS cm

-1

Correction of Ionization Errors: Figures 8.1-1 and 8.1-2 show the amount
of correction applied by the example program to compensate for ionization
effects on the measurements. Also shown is an ideal correction. Factors were

15