beautypg.com

2 edlog – Campbell Scientific CS547A-L Conductivity/Temperature Probe and A547A Interface User Manual

Page 13

background image

CS547A Conductivity and Temperature Probe and A547 Interface

'correct errors in the EC measurement due to temperature

Therm107

(TempDeg_C,1,3,Vx2,0,250,1,0)

C25mScm = (Ct * 100)/(((TempDeg_C-25) * TempCoef) + 100)


'end scan loop by calling output table

CallTable

ECSample


NextScan

EndProg

5.2.2 Edlog

*Table 1 Program
01: 5

Execution

Interval

(seconds)


;Make a preliminary measurement of resistance for autoranging.

1: Full Bridge (P6)
1: 1

Reps

2: 15

±2500 mV Fast Range

3: 1

DIFF

Channel

4: 1

Excite all reps w/Exchan 1

5: 2500

mV

Excitation

6: 1

Loc [ Rs ]

7: -.001

Mult

8: 1

Offset


2: BR Transform Rf [X/(1-X)] (P59)
1: 1

Reps

2: 1

Loc [ Rs ]

3: 1

Multiplier

(Rf)

;
;Test the initial measurement to make a more accurate measurement.
;

3: CASE (P93)
1: 1

Case Loc [ Rs ]


4: If Case Location < F (P83)

1: 1.8

F

2: 30

Then

Do


5: AC Half Bridge (P5)

1: 1

Reps

2:

15

±2500 mV Fast Range

3: 2

SE

Channel

4: 1

Excite all reps w/Exchan 1

5: 2500

mV

Excitation

6: 1

Loc [ Rs ]

7: 1.0

Mult

8: 0.0

Offset

7