B.1.6 set power operation mode, B.1.6 set, Power – Campbell Scientific CS475, CS476, and CS477 Radar Water Level Sensor User Manual
Page 43: Ndix b.1.6, set power operation mode

Appendix B. SDI-12 Commands/Changing Settings
The factory default water conditions are set to 1, which is smooth. The water 
condition settings should closely mimic the actual water conditions during 
normal river flow. 
Change this setting by first using the aXSWC+n! command (where n=0 
(custom), 1 (smooth), 2 (medium), or 3 (rough)) followed by the aXGWC! 
(Get Water Conditions) command.  TABLE B-5 shows an example of the 
command and response for changing this setting. 
TABLE B-5. Example for Setting Water Conditions
Initial Command
Response
0XSWC+2! 
Where, 
0—sensor’s address; 
2—the new water condition setting 
(2 = medium). 
00011
0—sensor’s address; 
001—the amount of time (in seconds) 
that you must wait before sending 
another command; 
1—the number of values that will be 
placed in the buffer. 
Subsequent Command
Response
0XGWC! 
Where zero is the sensor address. 
This is the send water conditions 
command. 
0+2
0—sensor’s address; 
2—the new water condition setting 
(2 = medium). 
B.1.6 Set Power Operation Mode
The following three power operation modes are available:
• 1 (ON—sensor is always on until a new set power command is received.)
• 2 (AUTO—sensor is powered by an incoming request from the SDI-12
bus and sends back a response including the information on power up 
time.) 
• 0 (OFF—this mode is typically not recommended; sensor is completely off
until a new set power command is received)
The OFF power mode should only be used by advanced 
users who want to turn the sensor off for extended time 
periods. In this mode, the sensor is completely off and 
only responds to a new set power command. 
CAUTION
The factory default Power Operation Mode is 1 (ON). In this mode of 
operation, the instrument is continuously making measurements and draws 
approximately 13.5 mA. The AUTO Power Operation Mode puts the 
instrument in quiescent mode between measurement request queries. 
B-5
