CS475, CS476, and CS477 Radar Water Level Sensor
CS475
CS476/CS477
1
2
FIGURE 4-1. Polarization markings (see TABLE 4-4 for label
descriptions)
TABLE 4-4. Description of Polarization Markings Labels
Sensor
Description
1 CS475
Polarization marks are designated by the mounting loop
screws.
2 CS476/CS477
Polarization mark is machine-tooled.
4.2 Step 2 — Do a False Echo Learn Command
The 25616, Adjustment/Display Module, or the terminal emulator
in LoggerNet or PC400 can be used to enter SDI-12 commands.
Appendix B.2, Using Terminal Emulator and a Datalogger to
Send Commands, describes entering commands using the
terminal emulator.
NOTE
Do a Start False Echo Learn command followed by a Send Data command
(see TABLE 4-4). To start false echo learn, do the aXSFEL+nnn.nnn!
command (where nnn.nnn = the actual distance to the water) followed by the
aD0! (Send Data) command. TABLE 4-5 shows an example of the command
and response.
Any echo occurring 0.5 m (1.6 ft) short of the distance you entered will be
considered noise.
Appendix B describes this command in further detail.