beautypg.com

Caution – Aalborg TPV User Manual

Page 16

background image

16

1. Remove the 4 thumbscrews from the transparent safety

cover, then lift the cover away from the pump face and set
it aside.

2. Loosen the turnbuckle rod between the top and bottom

parts of the tubing holder subassembly, until the
subassembly is movable.

3. Lift the tubing holder straight away from the pump face.
4. To load the tubing, begin to lay the tubing in the pump

channel, with both ends on the righthand side.

5. Manually turn the tri-star shaped head to turn the rollers as

needed so you can insert the tubing against the occlusion
wall.

6. When the tubing is in place, reinstall the tubing holder, with

the cutout on each end cupping the tubing. Make sure to
align the groove in the tubing holder end blocks with the
key ridge on the pump face.

7. Finger tighten the turnbuckle rod between the ends of the

tubing holder until the tubing is securely help in place, top
and bottom.

8. Hold the safety cover in place as you reinstall all 4

thumbscrews. Finger tighten the thumbscrews.

CAUTION:

NEVER RUN THE PUMP WITHOUT THE

SAFETY COVER IN PLACE.

This manual is related to the following products: