1. Remove the 4 thumbscrews from the transparent safety
cover, then lift the cover away from the pump face and set
it aside.
2. Loosen the turnbuckle rod between the top and bottom
parts of the tubing holder subassembly, until the
subassembly is movable.
3. Lift the tubing holder straight away from the pump face.
4. To load the tubing, begin to lay the tubing in the pump
channel, with both ends on the righthand side.
5. Manually turn the tri-star shaped head to turn the rollers as
needed so you can insert the tubing against the occlusion
wall.
6. When the tubing is in place, reinstall the tubing holder, with
the cutout on each end cupping the tubing. Make sure to
align the groove in the tubing holder end blocks with the
key ridge on the pump face.
7. Finger tighten the turnbuckle rod between the ends of the
tubing holder until the tubing is securely help in place, top
and bottom.
8. Hold the safety cover in place as you reinstall all 4
thumbscrews. Finger tighten the thumbscrews.
CAUTION:
NEVER RUN THE PUMP WITHOUT THE
SAFETY COVER IN PLACE.