Aalborg PWE Digital User Manual
Page 72

67
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Pilot
To
talizer
Sets and controls action of
the Pilot flow totalizer
.
NOTE:
Pilot T
o
talizer reading is
stored in SRAM (volatile)
memor
y. Power cycle will
reset Pilot T
o
talizer reading
to zero.
10
PT
Z (Reset to zero)
PTZ
F (start totalizer
at flow %F
.S.)
(flow %FS)
PTF:
L (Limit gas
volume in
current E.U.)
(gas volume)
PTL:
D (disable
totalizer)*
PT
:D
E (enable
totalizer)
PT
:E
R (read current
totalizer volume)
PTR:
(in current EU)
S (setting status)
PTS: Mode, Start, Limit
Low Flow
Cut Off
Display /Change Meter Low
Flow Cut Off settings in %F
.S.
11
CO
alue> (0 to 10.0%) CO: alue> Example: CO:2.0 No Argument (Returns Current Cut off V alue settings) CO: alue> Example: CO:2.0 Fluid Density Read and set Fluid Density in g/cm3 12 FD (New Density V alue) 0.01 Density 5.0 g/cm3 FD: No Argument (Returns Current Density in g/cm3) FD: Example: FD:1.000