User's Guide ______________________________________________________________________

26 __________________________________________________________________ M210669EN-D

* Defined as the ±2 standard deviation limits of end-point non-linearity, hysteresis
error or repeatability error.

** Defined as ±2 standard deviation limits of inaccuracy of the working standard
including traceability to NIST.

*** Defined as the root sum of the squares (RSS) of end-point non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room temperature.

**** Defined as ±2 standard deviation limits of temperature dependence over the
operating temperature range.

General

Supply voltage

with RS485

8 ... 18 VDC

with RS232C

5 ... 28 VDC

with RS232C/TTL

5 ... 28 VDC

reverse polarity protected

Current consumption

normal mode

< 15 mA (factory setting)

power down mode

< 0.8 mA

shutdown

mode

0.2

mA

Shutdown ON/OFF
Settling time at power up

2 s

Serial I/O (factory setting *)

RS232C

RS232C /TTL (optional)

RS485, non isolated (optional)

parity

none, even*, odd

data bits

7*, 8

stop bits

1*, 2

Baud rate

1200, 2400, 4800, 9600*, 19200

Response time

1 s*

Resolution

0.01 hPa (1 measurement/s)

0.05 hPa (20 measurements/s)

Max. pressure

5 000 hPa abs.

Pressure connector

M5 (10-32) internal thread

Pressure fitting

barbed fitting for 1/8" I.D. tubing

Housing
electronics

IP65

sensor

IP53

Housing material

PC plastic

Pressure connector

M5 (10-32) internal thread

Pressure fitting

barbed fitting for 1/8" I.D. tubing

Supply/output cable length

1, 2, 3, 5, or 10 m

Weight

Instrument

110

g

Cable

28

g/m