Output with or without the Unit
.FORM.0
The command cancels the outputting of the unit after pressure reading.
This is also the factory setting for the output format.
1017.61
.FORM.1
The command adds the unit to the outputting. The standard output
field is for 6 characters. The first character is always a space.
1017.61 hPa
NOTE
To activate the new setting, please remember to reset the barometer
with the command .
RESET
.RESET Resetting the Instrument
.RESET
The command resets the barometer and activates the setting changes.
Setting the Pressure Corrections
On/Off
These commands are used to set multi-point corrections either on or
off. See also the chapter Calibration and Adjustment on page 21.
.MPCON
The command activates the multi-point corrections.
.MPCOFF
The command deactivates the multi-point corrections.
NOTE
To activate the new setting, please remember to reset the barometer
with the command .
RESET