Chapter 6, Technical data, Technical specification – Vaisala PTB210 (serial) User Manual
Page 27: Operating range, Accuracy, Technical specification operating range
![background image](/manuals/408519/27/background.png)
Chapter 6 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 25
CHAPTER 6
TECHNICAL DATA
Technical Specification
Operating Range
Pressure range
500 ... 1100 hPa
50 ... 1100 hPa
Operating temperature range
-40 ... +60
C
Humidity range
non condensing
Accuracy
500…1100 hPa
Class A
ClassB
Non linearity*
0.10 hPa
0.15 hPa
Hysteresis*
0.05 hPa
0.05 hPa
Repeatability*
0.05 hPa
0.05 hPa
Calibration uncertainty**
± 0.07 hPa
± 0.15 hPa
Total accuracy (20
C)*** 0.15 hPa
0.20 hPa
Temperature dependence
(-40 ... +60
C)****
± 0.20 hPa
± 0.20 hPa
Total accuracy***
(-40 ... +60
C)
0.25 hPa
0.30 hPa
Long term stability
0.10 hPa/year 0.10 hPa/year
50…1100 hPa
Non linearity*
0.20 hPa
Hysteresis*
0.10 hPa
Repeatability*
0.10 hPa
Calibration uncertainty**
± 0.20 hPa
Total accuracy (20
C)***
0.35 hPa
Temperature dependence
(-40 ... +60
C)****
± 0.40 hPa
Total accuracy***
(-40 ... +60
C) 0.50 hPa
Long term stability
± 0.20 hPa/year