Self-testing, Basic information of the barometer settings – Vaisala PTB210 (serial) User Manual

Page 22

background image

User's Guide ______________________________________________________________________

20 __________________________________________________________________ M210669EN-D

.RON

The .

RON

command sets the terminating resistor on.

.ROFF

The .

ROFF

command sets the terminating resistor off.

Self-Testing

.? Basic Information of the
Barometer Settings

.?

The command lists the basic information of the barometer settings.
The following example shows also the factory settings for 50…1100
hPa model.

.? (command not visible)
PTB210 Ver 1.0

CAL DATE :1999-06-10
ID CODE :10
SERIAL NUMBER :T00100004
MULTIPOINT CORR:ON
MEAS PER MINUTE: 60

AVERAGING : 0
PRESSURE UNIT : hPa
Pressure Min...Max: 500 1100 (50 1100 for model 50…1100)

LOW CURRENT MODE

RS485 RESISTOR OFF