background image

Chapter 9 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 79

General

IECEx (VTT)

Ex ia IIC T4 Ga
IECEx VTT 09.0002x issue No:1

Safety factors

U

i

= 28 V, I

i

= 100 mA, P

i

= 700 mW

C

i

= 1 nF, L

i

negligibly low

Environmental specifications

T

amb

= -40 ... +60 °C (-40 ... +140 °F)

P

amb

= 0.8 ... 1.1 bar

Table 12

General Specifications

Property

Description / Value

Operating voltage
with serial port (service mode)

12 ... 28 V
15 ... 28 V

Connections

screw terminals,

0.33 ... 2.0 mm

2

wires

(AWG 14-22)

Cable bushings

M20×1.5 (7.5 ... 12 mm)
M20×1.5 (10.5 ... 15 mm)

Conduit fitting

M20×1.5 / NPT 1/2"

Operating temperature range for electronics
with display

Storage temperature range

-20 ... +60 °C

-40 ... +70 °C

Housing material

G-AlSi 10 Mg (DIN 1725)

Housing classification

IP 66 (NEMA 4X)

Housing dimensions

164 Ч 115 Ч 62 mm

Housing weight

950 g

Fully electromagnetically compatible
according to standards

EN 61326-1: Electrical
equipment for measurement,
control and laboratory use -
EMC requirements; Industrial
environment

NOTE

IEC 1000-4-5 complies only when using external EXi approved surge
arrester on safe area.