Adjustment commands, Ct temperature adjustment – Vaisala HMT360N User Manual
Page 57
Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 55
Example:
>std -50 50
Td lo: -50.00 %RH
Td hi: 50.00 %RH
>
Adjustment Commands
Turn the internal DIP switch Calibration Enabled/Disabled to
position ON before the adjustment and return it to position OFF after
the adjustment.
CT Temperature Adjustment
Syntax: CT
The transmitter asks and measures temperature readings and calculates
the calibration coefficients.
Example:
>ct
T : 0.811 1. ref ? 0.5
Press any key when ready ...
T : 99.122 2. ref ? 99.5
OK
>
The OK indicates that the adjustment was succesful. When performing
one-point adjustment press only Enter for the second reference.
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)