Send outputting measurement values, R activating continuous output, S stopping continuous output – Vaisala HMT360N User Manual
Page 59: Intv setting output interval

Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 57
SEND Outputting Measurement Values
Syntax: SEND
This command outputs the measured values in one point.
R Activating Continuous Output
Syntax: R
With the command R the transmitter outputs measured values
continuously.
S Stopping Continuous Output
Syntax: S
The continuous outputting is stopped with the command S. If outputting
is active, this command is not echoed.
INTV Setting Output Interval
Syntax: INTV n xxx
Sets the output interval when the transmitter outputs measurement
readings. The time interval is used when the continuous output is active.
Example, the output interval is set to 10 minutes:
>intv 10 min
Output interval: 10 MIN
>
where
n
=
1 ... 255
xxx =
S, MIN or H