Scaling analog outputs – Vaisala HMT360N User Manual
Page 53

Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 51
0505-289
Press button C to exit the display command mode or continue by setting
the pressure.
Scaling Analog Outputs
Turn the internal Calibration Enabled/Disabled DIP switch of the
transmitter to upward position (on). Press button C on the display cover
and a text similar to the following appears:
0604-053
Numbers on the second line indicate the low end scaling of Ch 1
currently stored in the transmitter memory. The text "Set Lo" on the
lower left corner indicates that you can now change the low end scaling
with buttons Up and Dn. Acknowledge the value with button E and a
text similar to the following appears:
0604-054
Numbers on the second line indicate the high end scaling of Ch 1. The
text "Set Hi" on the lower left-hand corner indicates that you can now
change the high end scaling with buttons Up and Dn. Acknowledge
with button E.
NOTE
Remember to restore the DIP switch settings.