S scaling analog outputs – Vaisala HMT360N User Manual
Page 56

User's Guide ______________________________________________________________________
54 __________________________________________________________________ M210744EN-C
Turn the internal DIP switch Calibration Enabled/Disabled to
position ON before the selecting the analog outputs and return it to
position OFF after making the selection. Output quantities and their
abbreviations are listed in the table below:
Example:
>asel td t
Ch1 Td lo: -40.00 ’C
Ch1 Td hi: 100.00 ’C
Ch2 T lo: -40.00 ’C
Ch2 T hi: 100.00 ’C
>
S Scaling Analog Outputs
Syntax: Szz aa.a bb.b
Turn the internal DIP switch Calibration Enabled/Disabled to
position ON before the scaling the analog outputs and return it to
position OFF after scaling.
Quantity
Abbreviation
temperature
T
dewpoint temperature
Td
volume concentration
ppmv
water content
w (when scaling water content, the
abbreviation is "ng1", see note
below)
NOTE
When scaling water content w (mg/m
3
or lb/mmscf) with the natural
gas transmitter, issue command SNG1.
where
zz =
Quantity (T, Td, ppm, w)
aa.a =
Lower limit of the quantity
bb.b=
Upper limit of the quantity