beautypg.com

2 hmp42 probe, 1 relative humidity, 2 temperature – Vaisala HMP42 User Manual

Page 50: Hmp42 probe

background image

HMI41 AND HMP42
Operating Manual

M210317EN-A

46

11.2

HMP42 probe

11.2.1

Relative humidity

Measurement range

0...100 %RH non-condensing

Accuracy (at +20 °C)

when calibrated against
salt solutions (ASTM E104-85):

±2 %RH (0...90 %RH)

±3 %RH (90...100 %RH)

Temperature dependence
of electronics

±0.05 %RH/

°C

Typical long-term stability

better than 1 %RH per year

Response time (90%) at 20

°C

in still air with

30 s

Humidity sensor

HUMICAP

®

MINI

11.2.2

Temperature

Measurement range
(sensor head)

-40...+100

°C

Accuracy at +20 °C

±0.2 °C

Accuracy:

-40

-20

0

20

40

60

80

100

0.2

0.4

0.6

-0.2

-0.4

-0.6

°C

0

Temperature sensor

Pt 100 (IEC 751 1/3 Class B)

This manual is related to the following products: