beautypg.com

Setup procedure, 1 entering the setup mode, Entering the setup mode – Vaisala HMP42 User Manual

Page 41

background image

HMI41 AND HMP42

M210317EN-A

Operating Manual

37

8.

SETUP PROCEDURE

The indicator settings can be changed in the

SETUP

mode. Note that manual

setup procedure is necessary if some of the factory settings have to be changed
and if the indicator or the probe does not have letters ID in the instrument
label.

The factory settings of the HMI41 are the following:

• display units:

0 (metric)

• automatic power-off:

5 minutes

• display quantities:

0

(= RH, T, Td)

• pressure:

1013.25 hPa

(1 hPa = 1 mbar)

(for wet bulb temperature and mixing ratio calculations)

• probe type

AUT

(or 1, see below)

• start

1

Indicators marked with letters ID have the automatic probe recognition as
default (

AUT PROBE

) and previous versions the probe type 1. If the indicator

does not recognize the probe type automatically, set the type manually to
2 when using the HMP42.
Note also that the HMP42 probe cannot be used
with the HMI41 unless the software version is 1.02 or greater. To check the
version, turn the HMI41 on with the ON/OFF button. The version appears on
the display in a couple of seconds if it is 1.02 or greater. If it does not appear,
contact Vaisala or a Vaisala representative for further information.

8.1

Entering the setup mode

For entering the setup mode, first press the ON/OFF button and the following
appears:

RH

set min max

hyst hold

Lo bat Hi

P

Pa

h

g/m

T

Td

°C °F

°C °F

%

Then release the ON/OFF button and within 1 - 2 seconds press simultane-
ously both ENTER and MODE buttons until the following text appears on the
display:

This manual is related to the following products: