Posey Soft Belt User Manual
Application instructions, Rx only soft belt
Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com
Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014
MDSS GmbH
Schiffgraben 41
D-30175 Hannover, Germany
EC REP
APPLICATION
INSTRUCTIONS
BED SAFETY
• ALWAYS use Hospital Bed Safety Workgroup
(HBSW) (http://www.fda.gov/cdrh/beds/
modguide.html) compliant side rails in the UP
position and fill ALL gaps to reduce the risk of
entrapment.
• Use side rail covers and gap protectors to help prevent the patient’s body
from going under, around, through or between the side rails. A failure to
do so may result in serious injury or death if a patient becomes suspended
or entrapped. Posey offers a full range of side rail pads and/or gap
protectors to cover gaps.
INDICATIONS FOR USE:
CONTRAINDICATIONS:
ADVERSE REACTIONS
• Severe emotional, psychological, or physical problems may occur: if the applied device is uncomfortable; or if it severely
limits movement. If symptoms of these problems ever appear for any reason, get help from a qualified medical
authority and find a less restrictive, product or intervention.
•
NEVER use a 5th point restraint on a patient:
•
With a pelvic fracture, supra-pubic catheter, ostomy; percutaneously placed feeding tube or recent incision.
•
With a history of cardiac or pulmonary disorders, or thoracic fractures.
•
To restrain head or neck.
• Patients who need a supplemental restraint (5th point) of the thighs, pelvis, or chest, and
who are already restrained at all four extremities.
• Patients requiring a positioning device to assist medical treatment.
• Patients assessed to be at risk of injury from a fall.
•
DO NOT use on patients with: ostomy, colostomy, or G-tubes; hernias, severe Cardio Obstructive Pulmonary Disease
(COPD); or with post-surgery tubes, incisions or monitoring lines. These could be disrupted by a restraint.
•
DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or
suicidal, except as a 5th point restraint (see below).
APPLICATION INSTRUCTIONS:
Before use, check device for damage. Discard if you have any
questions about patient safety.
Applicable Products:
REF
4125, 4125L, 4125Q
DESCRIPTION OF PRODUCT: Foam padded belt with cotton straps. For chair or hospital bed use.
Rx ONLY
SOFT BELT
© 2009 Posey Company. All rights reserved.
I9205 040609
4125
POSEY SOFT BELT
REF
4125 Bed & chair; 4½”W x 16½” (11
cm x 42 cm) belt pad w/6 foot
(2 m) straps
REF
4125Q Bed & chair; 4½”W x 16½” (11
cm x 42 cm) belt pad, 6 foot
(2 m) straps w/quick-release
buckles
REF
4125L Bed & chair; 4½”W x 30” (11
cm x 76 cm) belt pad w/8 foot
(2 2/5 m) straps
APPLICATION INSTRUCTIONS: CHAIR
1. Position the patient as far back in the seat as possible,
with the buttocks against the back of the chair.
2. Lay the lap belt across the patient’s thighs with the
foam facing in.
3. Bring the ends of the connecting straps down at a
45-degree angle between the seat and the wheelchair
sides (fig. 1). Criss-cross the straps behind the chair
and draw them around the opposite side kick spurs.
4. Secure each connecting strap, out of the patient’s
reach, using a quick-release tie or buckle (figs. 2A
and 2B). Check that the straps are secure and will
not change position, loosen, or tighten if the patient
pulls on them, or if the chair is adjusted.
5. The belt must be snug, but not interfere with
breathing. To check for proper fit, slide an open hand
(flat) between the belt and the patient.
APPLICATION INSTRUCTIONS: HOSPITAL BED
1. Bring the belt around the patient’s waist with the foam pad facing in.
2. Criss-cross the straps behind the patient, and feed each strap through the
positioning loops on the ends of the blue foam pad. The belt must be snug,
but not interfere with breathing. To check for proper fit, slide an open hand
(flat) between the belt and the patient.
3. Secure each connecting strap around a movable part of the bed frame, at
waist level, using a quick-release tie or buckle. Check that the straps are
secure and will not change position, loosen, or tighten if the patient pulls on
them, or if the bed is adjusted.
APPLICATION INSTRUCTIONS: 5TH POINT RESTRAINT
1. Bring the belt around the patient’s waist, chest, or legs with the foam pad
facing in. The belt must be snug, but not interfere with breathing. To check
for proper fit, slide an open hand (flat) between the belt and the patient.
2. Secure each connecting strap around a movable part of the bed frame, using
quick-release ties or buckles. Check that the straps are secure and will not change
position, loosen, or tighten if the patient pulls on them, or if the bed is adjusted.
NOTE: When using over the chest, the straps should go under the arms and
attach at chest level, to a movable part of the bed frame. This will prevent the
patient from sliding down and becoming entangled.
FIG. 2A
FIG. 1
45˚
ADDITIONAL SAFETY AND LAUNDERING
INSTRUCTIONS ON OTHER SIDE
FIG. 2B
FIG. 4
FIG. 5
FIG. 3
• Monitor per facility policy to ensure that the patient cannot slide down, or
fall off the chair seat or mattress and become suspended or entrapped
(figs. 3, 4, and 5).
• There is a risk of chest compression or suffocation if the patient’s body
weight is suspended off the mattress or chair seat. Use extreme caution
with chair cushions. If a cushion dislodges, straps may loosen and allow
the patient to slide off the seat (figs. 3, 4, and 5).
• STOP USE AT ONCE: if the patient has a tendency to slide forward or
down in the device; or is able to self-release.
NOTE: A restraint with a pelvic piece will reduce the risk of sliding, or of
the patient pulling the device over his or her head. See Posey Catalog.
• NEVER leave a patient requiring a 5th point restraint alone. One-on-one
supervision is required in case of an emergency.
Heed these warnings to reduce the risk of serious injury or death:
Make sure patient wears proper undergarments to protect skin.