MONITOR PER FACILITY POLICY. Check to ensure that:

• Straps cannot slide in any direction or loosen if the patient

pulls on them, or if the bed is adjusted;

• Cuffs are properly secured. If applied too tightly,

circulation will be restricted; if applied too loosely, the

patient may be able to slip his or her limb from the device;

• Cuffs are attached in a way that the patient is not able to

use his or her teeth or otherwise remove the device;

• Cuffs are intact, and not torn or damaged. DO NOT allow

patients to ingest product material.

ADDITIONAL OR DIFFERENT BODY OR LIMB RESTRAINTS

MAY BE NEEDED:

• If the patient pulls violently against the bed straps;

• To reduce the risk of the patient getting access to the line/

wound/tube site;

• To prevent the patient from flailing or bucking up and

down causing self-injury.

See Posey Catalog.

BED SAFETY

• ALWAYS use Hospital Bed Safety

Workgroup (HBSW) (http://www.fda.

gov/cdrh/beds/modguide.html)

compliant side rails in the UP position

and fill ALL gaps to reduce the risk of

entrapment.

• Use side rail covers and gap protectors to help prevent

the patient’s body from going under, around, through or

between the side rails. A failure to do so may result in

serious injury or death if a patient becomes suspended

or entrapped. Posey offers a full range of side rail pads

and/or gap

protectors to

cover gaps.

1. Use method a. or b. below to attach straps (repeat steps

1-6 on each side):

a. Triangulation process; to restrict

patient’s range of motion:

Separate the straps and attach

them at different points along

a movable part of the bed

frame, out of the patient’s reach,

using quick-release ties (fig. 1).

b. To increase patient’s range of motion: Place the straps

together and attach to a single point along a movable

part of the bed frame, out of the patient’s reach, using

quick-release ties.

2. Wrap the limb holder cuff around

the patient’s wrist so the buckle and

connecting strap is on the ulnar side

of the wrist (opposite the thumb)

(fig. 2).

3. Secure the hook and loop fastener.

Slide ONE finger (flat) between the

cuff and the inside of the patient’s

wrist to ensure proper fit (fig. 3).

The strap must be snug, but not

compromise circulation.

4. Close the quick-release buckle on

the cuff. Insert TWO fingers (flat)

under the buckle and pull the strap

snug, but not so tight as to restrict

circulation (fig. 4).

5. Release the quick-release buckle,

twist, and reconnect (fig. 5). Listen

for a “snapping” sound. Pull firmly on

straps to ensure a good connection.

6. Adjust the bed strap(s) to allow

desired freedom of movement,

without compromising patient or

caregiver safety.

Applicable Products:

REF

2541

DEscription oF proDuct: Limb holder for limiting limb movement. For bed use only.

rx onlY

liMB HolDErs

© 2009 Posey Company. All rights reserved.

I9283 062209

2541

posEY liMB HolDErs

REF

2541 Double straps with tie

ends and quick-release

buckle on the cuff strap.

ApplicAtion instructions:

Fig. 2

Fig. 3

Fig. 1

Fig. 4

ADDitionAl sAFEtY AnD lAunDEring

instructions on otHEr siDE

Fig. 5

inDicAtions For usE:

• Patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling).
• Patients assessed to be at risk of line pulling, which may prevent monitoring of vital signs.
• Patients whose picking, pulling, scratching, or peeling exacerbates a skin condition, causes self-

injury, or compromises wound site integrity.

ADVErsE rEActions

• Severe emotional, psychological, or physical problems may occur: if the applied device is

uncomfortable; or if it severely limits movement. If symptoms of these problems ever appear

for any reason, get help from a qualified medical authority and find a less restrictive, product or

intervention.

contrAinDicAtions:

DO NOT use this device on a patient who is or becomes: suicidal; highly aggressive or

combative; self-destructive; or deemed to be an immediate risk to others, UNLESS the patient

is under constant supervision.

Before use, check device for damage. Discard if you

have any questions about patient safety.

NOTE: To remove cuffs, unsnap quick-release buckles and

release hook and loop fasteners.

NEVER use on a patient:

With a dislocation or fracture on the restrained limb; or

If an IV or wound site could be compromised by the device.

See the Posey Catalog for other options for such a patient.