beautypg.com

Posey Secure Straps User Manual

Mitt connecting strap, Application instructions, Rx only

background image

Application

Instructions

DESCRIPTION OF PRODUCT:

Connecting straps (for hospital bed use only).

Rx ONLY

Mitt Connecting Strap

© 2011 Posey Company. All rights reserved.

I92107 REV. A 101711

AddItIONAL SAfetY ANd LAuNdeRINg

INStRuCtIONS ON OtheR SIde

2809

Application Instructions

:

Posey Mitt Connecting Strap

REF

2809

Connecting Strap, 48” (123 cm)

1. Apply a Posey Mitt to the patient’s hand following the specific mitt

instructions for use.

2. Wrap the strap around the patient’s wrist or pass it through the

loops on the mitt.

3. Use Posey Quick-Release Tie (see drawings on reverse) to secure the

end of the strap to a movable part of the bed frame. Tie strap at

a point midway between the patient’s wrist and elbow, out of the

patient’s reach.

BED SAFETY
• ALWAYS use Hospital Bed Safety

Workgroup (HBSW) (http://www.fda.

gov search keyword “HBSW”) compliant

side rails in the UP position and fill ALL

gaps to reduce the risk of entrapment.

• Use side rail covers and gap protectors to help prevent the

patient’s body from going under, around, through or between

the side rails. A failure to do so may result in serious injury or

death if a patient becomes suspended or entrapped. Posey

offers a full range of side rail pads and/or gap protectors to

cover gaps.

Indications for use:

• Patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling).
• Patients assessed to be at risk of line pulling, which may prevent monitoring of vital signs.
• Patients whose picking, pulling, scratching, or peeling exacerbates a skin condition, causes

self-injury, or compromises wound site integrity.

NEVER use mitts on a patient:

If an IV or wound site could be compromised by the device; or

With a dislocation or fracture on the affected limb.

See the Posey Catalog for other options for such a patient.

Contraindications:

DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or suicidal.

Adverse Reactions

• Severe emotional, psychological, or physical problems may occur: if the applied device is

uncomfortable; or if it severely limits movement. If symptoms of these problems ever appear for any
reason, get help from a qualified medical authority and find a less restrictive, product or intervention.

MONITOR PER FACILITY POLICY. Check to ensure that:
• Connecting straps cannot slide in any direction or loosen if the

patient pulls on them, or if the bed is adjusted;

• Mitts and straps are properly secured. If applied too tightly,

circulation will be restricted; if applied too loosely, the patient

may be able to slip his or her limb from the device;

• The patient cannot use his or her teeth or otherwise remove

the device and inflict self-injury;

ADDITIONAL OR DIFFERENT BODY OR LIMB RESTRAINTS MAY BE

NEEDED (See Posey Catalog):
• If the patient pulls violently against the bed straps;
• To reduce the risk of the patient getting access to the line/

wound/tube site;

• To prevent the patient from flailing or bucking up and down and

causing self-injury.

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

MDSS GmbH

Schiffgraben 41

D-30175 Hannover, Germany

EC REP