beautypg.com

Posey Pediatric Limb Holders User Manual

Application instructions, Rx only pediatric limb holders

background image

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

MDSS GmbH

Schiffgraben 41

D-30175 Hannover, Germany

EC REP

ApplicAtion

instructions

1. Wrap the limb holder cuff around the patient’s wrist or

ankle with the blue side toward the skin.

2. Secure the hook and loop fastener. Slide one finger (flat)

between the cuff and the inside of the patient’s wrist or

ankle to ensure proper fit. The cuff must be snug, but not

compromise circulation.

3. Take the strap directly away from the patient and secure

out of the patient’s reach.

NOTE: DO NOT wrap the straps around the patient’s limb.
4. Leave straps loose enough for extended range of motion,

but not enough to allow self-injury or disruption of lines or

extubation.

ApplicAtion

instructions

Applicable Products:

REF

4733, 4734

DEscription oF proDuct: Limb holder for limiting limb movement of infant and pediatric patients.

rx onlY

pEDiAtric liMB HolDErs

© 2009 Posey Company. All rights reserved.

I92102 REV A 111309

4733

posEY pEDiAtric liMB HolDErs

REF

4733 Infant to 2 years

REF

4734 2-5 years

ApplicAtion instructions:

inDicAtions For usE:
• Patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube

pulling).

• Patients assessed to be at risk of line pulling, which may prevent monitoring of vital

signs.

• Patients whose picking, pulling, scratching, or peeling exacerbates a skin condition,

causes self-injury, or compromises wound site integrity.

ADVErsE rEActions
• Severe emotional, psychological, or physical problems may occur: if the applied device

is uncomfortable; or if it severely limits movement. If symptoms of these problems

ever appear for any reason, get help from a qualified medical authority and find a less

restrictive, product or intervention.

contrAinDicAtions:

NEVER use on a patient:

With a dislocation or fracture on the restrained limb; or

If an IV or wound site could be compromised by the device.

See the Posey Catalog for other options for such a patient.

Before use, check device for damage. Discard

device if you have any questions about patient safety.

MONITOR PER FACILITY POLICY. Check to ensure that:

• Straps cannot slide in any direction or loosen if the patient

pulls on them, or if the bed is adjusted;

• Cuffs are properly secured. If applied too tightly,

circulation will be restricted; if applied too loosely, the

patient may be able to slip his or her limb from the device;

• Cuffs are attached in a way that the patient is not able to

use his or her teeth or otherwise remove the device;

• Cuffs are intact, and not torn or damaged. DO NOT allow

patients to ingest product material.

ADDITIONAL OR DIFFERENT BODY OR LIMB RESTRAINTS

MAY BE NEEDED:

• If the patient pulls violently against the bed straps;

• To reduce the risk of the patient getting access to the line/

wound/tube site;

• To prevent the patient from flailing or bucking up and

down causing self-injury.

See Posey Catalog.

BED SAFETY
• ALWAYS use Hospital Bed Safety

Workgroup (HBSW) (http://www.fda.

gov/cdrh/beds/modguide.html)

compliant side rails in the UP position

and fill ALL gaps to reduce the risk of entrapment.

• Use side rail covers and gap protectors to help prevent the

patient’s body from going under, around, through or

between the side rails. A failure to do so may result in

serious injury or death if a patient becomes suspended or

entrapped. Posey offers a full range of side rail pads and/

or gap protectors to cover gaps.

ADDitionAl sAFEtY AnD lAunDEring

instructions on otHEr siDE